[Home] ]
Data Verification Summary Report
Package 32926
Data Verifiers: Michelle Wolfe & Tammy Chang
Introduction
The following data verification summary report covers environmental soil samples and associated field quality control (QC) samples collected from the Camp Stanley (under RL83) on June 14, 2000. Samples in the following laboratory Sample Delivery Group (SDG) were analyzed for semivolatile organic compounds (SVOCs); volatile organic compounds (VOCs); and metals including barium, chromium, copper, nickel, zinc, arsenic, cadmium, lead and mercury:
32926 |
Field quality control samples collected were trip blank; equipment blank; matrix spike/matrix spike duplicates (MS/MSD); and field duplicates. During the initiation of this project, it was determined that ambient blanks were not necessary due to the absence of a source at the site. The trip blanks were analyzed for volatile organics only. All other field quality control samples were analyzed for the same parameters as their associated samples.
All samples were collected by Parsons Engineering Science (Parsons ES). All analyses were performed by APPL, Inc. following procedures outlined in the AFCEE QAPP, version 3.0.
Evaluation Criteria
The data submitted by the laboratory has been reviewed and verified following the guidelines outlined in the AFCEE QAPP, version 3.0. Information reviewed in the data packages include sample results; the summary of laboratory quality control results; case narrative; raw data; and chain-of-custody forms. The analyses and findings presented in this report are based on the reviewed information, and whether guidelines in the AFCEE QAPP were met.
SVOC SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for semivolatile organic compounds (SVOCs).
SVOC analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 8270C. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples; LCS samples; and surrogate spikes. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
All MS/MSD %Rs were within acceptance criteria except for as follows:
Sample AOC 48-SS03 (0-0.5�) | |||
Analyte | MS %R | MSD %R | QC |
Benzoic Acid | 11.8 | 10.6 | 25-172 |
The benzoic acid results in the samples from the same site with similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.
The LCS and surrogate %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
All MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
The results for samples in this SDG were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP; | |
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All instrument performance check criteria was met. | |
All initial calibration criteria were met. | |
All continuing calibration criteria were met. | |
All second source verification criteria were met. | |
All internal standard criteria were met. |
There were two method blanks and one equipment blank associated with the SVOC analyses in this SDG. The method blanks were free of SVOCs above the RL. The equipment blank, RL83-EB09 contained 34.00 mg/L of naphthalene and 14.00 mg/L of phenol. No action was needed since the associated samples did not contain either of these analytes.
VOC SDG 32926
General
This SDG consisted of eleven (11) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples, one equipment blank sample and one trip blank sample. The samples were collected on June 14, 2000 and analyzed for volatile organic compounds (VOCs).
VOC analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 8260B. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples; LCS samples; and surrogate spikes. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
All MS/MSD %Rs were within acceptance criteria except for as follows:
Sample AOC 48-SS03 (0-0.5�) | |||
Analyte | MS %R | MSD %R | QC |
1,1,1,2- tetrachloroethane | 33.8 | - | 62-108 |
1,1,2,2-tetrachloroethane | 26.5 | - | 64-135 |
1,1,2-TCA | 32.4 | - | 65-135 |
1,1-DCA | 58.8 | 50.5 | 62-135 |
1,2,3-trichlorobenzene | 7.4 | 36.8 | 65-147 |
1,2,3-trichloropropane | 27.9 | - | 65-135 |
1,2,4-trichlorobenzene | 10.3 | 45.6 | 65-145 |
1,2,4-trimethylbenzene | 27.9 | - | 65-135 |
1,2-DCA | 38.2 | - | 58-137 |
1,2-DCB | 16.2 | - | 65-135 |
1,2-dibromo-3-chloropropane | 20.6 | - | 49-135 |
1,2-dichloropropane | 44.1 | - | 60-135 |
1,2-EDB | 14.7 | 58.8 | 65-135 |
1,3,5-trimethylbenzene | 36.8 | - | 62-135 |
1,3-DCB | 20.6 | - | 65-135 |
1,3-dichoropropane | 32.4 | - | 65-135 |
1,4-DCB | 17.6 | - | 65-135 |
1-chlorohexane | 51.5 | - | 65-135 |
2-chlorotoluene | 29.4 | - | 63-135 |
4-chlorotoluene | 23.5 | - | 64-135 |
benzene | 51.5 | - | 65-135 |
bromobenzene | 22.1 | - | 65-135 |
bromochloromethane | 44.1 | - | 63-135 |
bromodichloromethane | 32.4 | 60.3 | 65-135 |
bromoform | 14.7 | 33.8 | 65-135 |
bromomethane | 5.9 | 32.4 | 62-135 |
chlorobenzene | 32.4 | - | 65-135 |
chloroform | 50.0 | - | 64-135 |
chloromethane | 42.6 | - | 65-135 |
cis-1,2-DCE | 45.6 | - | 65-135 |
cis-1,3-dichloropropene | 1.5 | 22.1 | 64-135 |
dibromochloromethane | 23.5 | 54.4 | 63-135 |
dibromomethane | 39.7 | - | 59-137 |
ethylbenzene | 44.1 | - | 65-135 |
hexachlorobutadiene | 17.6 | 58.8 | 65-135 |
isopropylbenzene | 45.6 | - | 65-135 |
m&p-xylene | 41.2 | - | 65-135 |
methylene chloride | 50.0 | - | 65-135 |
n-butylbenzene | 26.5 | - | 65-135 |
n-propylbenzene | 39.7 | - | 65-135 |
naphthalene | 7.4 | 33.8 | 65-135 |
o-xylene | 36.8 | - | 65-135 |
p-isopropyltoluene | 23.5 | - | 65-135 |
sec-butylbenzene | 39.7 | - | 65-135 |
styrene | 23.5 | - | 65-135 |
TCE | 54.4 | - | 61-135 |
tert-butylbenzene | 38.2 | - | 65-135 |
tetrachloroethene | 55.9 | - | 61-135 |
toluene | 45.6 | - | 64-135 |
trans-1,2-DCE | 58.8 | - | 65-135 |
trans-1,3-dichloropropene | 4.4 | 26.5 | 56-135 |
- The %R was compliant. |
The MS/MSD were reanalyzed with similar results. The raw data package contains results from both injections. The results for the non-compliant analytes in the samples from the same site and with similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.
The LCS and surrogate %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
All MS/MSD RPDs were within acceptance criteria except for as follows:
Sample AOC 48-SS03 (0-0.5�) | ||
Analyte | RPD | QC |
1,1,1,2- tetrachloroethane | 82.1 | 30 |
1,1,1-TCA | 40.0 | 30 |
1,1,2,2-tetrachloroethane | 119.1 | 30 |
1,1,2-TCA | 85.7 | 30 |
1,1-DCA | 50.5 | 30 |
1,1-DCE | 38.7 | 30 |
1,1-dichloropropene | 36.4 | 30 |
1,2,3-dichlorobenzene | 133.3 | 30 |
1,2,3-trichloropropane | 118.3 | 30 |
1,2,4-trichlorobenzene | 126.3 | 30 |
1,2,4-trimethylbenzene | 111.6 | 30 |
1,2-DCA | 77.6 | 30 |
1,2-DCB | 122.8 | 30 |
1,2-dibromo-3-chloropropane | 111.1 | 30 |
1,2-dichloropropane | 66.7 | 30 |
1,2-EDB | 120.0 | 30 |
1,3,5-trimethylbenzene | 99.0 | 30 |
1,3-DCB | 121.1 | 30 |
1,3-dichoropropane | 92.7 | 30 |
1,4-DCB | 128.4 | 30 |
1-chlorohexane | 52.6 | 30 |
2,2-dichloropropane | 34.2 | 30 |
2-chlorotoluene | 110.1 | 30 |
4-chlorotoluene | 119.0 | 30 |
benzene | 54.2 | 30 |
bromobenzene | 120.0 | 30 |
bromochloromethane | 81.2 | 30 |
bromodichloromethane | 60.3 | 30 |
bromoform | 78.8 | 30 |
bromomethane | 138.5 | 30 |
chlorobenzene | 87.2 | 30 |
chloroethane | 39.4 | 30 |
chloroform | 62.6 | 30 |
chloromethane | 58.5 | 30 |
cis-1,2-DCE | 63.7 | 30 |
cis-1,3-dichloropropene | 175.0 | 30 |
dibromochloromethane | 79.2 | 30 |
dibromomethane | 89.8 | 30 |
dichlorodifluoromethane | 30.6 | 30 |
ethylbenzene | 71.0 | 30 |
hexachlorobutadiene | 107.7 | 30 |
isopropylbenzene | 94.0 | 30 |
m&p-xylene | 72.7 | 30 |
methylene chloride | 74.1 | 30 |
n-butylbenzene | 113.3 | 30 |
n-propylbenzene | 97.1 | 30 |
naphthalene | 128.6 | 30 |
o-xylene | 78.0 | 30 |
p-isopropyltoluene | 114.7 | 30 |
sec-butylbenzene | 94.1 | 30 |
styrene | 103.0 | 30 |
TCE | 49.0 | 30 |
tert-butylbenzene | 97.0 | 30 |
tetrachloroethene | 57.9 | 30 |
toluene | 62.2 | 30 |
trans-1,2-DCE | 46.2 | 30 |
trans-1,2-DCE | 142.9 | 30 |
trichlorofluoromethane | 32.4 | 30 |
vinyl chloride | 33.8 | 30 |
The results for the non-compliant analytes in the samples from the same site and with similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.
The field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
The results for samples in this SDG were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP; | |
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All instrument performance check criteria was met. | |
All initial calibration criteria were met. | |
All continuing calibration criteria were met. | |
All second source verification criteria were met. | |
All internal standard criteria were met for the continuing calibrations. |
There was one sample (AOC47-SS01) that had non-compliant internal standards (1,4-dichlorobenzene-D). The SW846 Method 8260B (section 7.4.7) specifies that the continuing calibration internal standard areas be compared to the initial calibration internal standard data. However, there is no mention of checking the samples internal standard areas. Therefore no action was taken for the sample with non-compliant internal standards.
There were four method blanks, one trip blank and one equipment blank associated with the VOC analyses in this SDG. The blanks were free of VOCs above the RL.
Metals SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for metals; barium, chromium, copper, nickel, and zinc.
The barium, chromium, copper, nickel, and zinc analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 6010B. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
All MS/MSD and LCS %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
All MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP; | |
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All initial and continuing calibration criteria were met. | |
All second source calibration criteria were met. | |
All interference check criteria were met. | |
All dilution test criteria were met except for as follows: |
Sample AOC 48-SS03 (0-0.5�) | ||
Analyte | %D | QC |
Barium Nickel Zinc | 13.6 20.4 11.7 | 10 10 10 |
The barium, nickel and zinc results in the associated samples were considered estimated and flagged �J�. The five-fold diluted chromium results were less than the reporting limit. Therefore, the dilution test was not applicable for the chromium.
All post digestion spike addition criteria were met. |
There were two method blanks, one equipment blank and several calibration blanks associated with the metal analyses in this SDG. All blanks were free of any metals above the RL.
Mercury SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for mercury.
The mercury analyses were performed using USEPA SW846 Method 7470A. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
The MS/MSD and LCS %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
All MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
All results were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP; | |
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All initial and continuing calibration criteria were met. | |
All second source calibration criteria were met. |
There were two method blanks and one equipment blank and several calibration blanks associated with the mercury analyses in this SDG. All the blanks were free of any mercury above the RL.
Arsenic SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for arsenic.
The arsenic analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7060A. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
The MS/MSD and LCS %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
The MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
All the results were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP; | |
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All initial and continuing calibration criteria were met. | |
All second source calibration criteria were met. | |
The five-fold diluted arsenic result was less than the reporting limit. Therefore, the dilution test was not applicable. | |
The recovery test criteria was met. |
There were two method blanks, one equipment blank and several calibration blanks associated with the arsenic analyses in this SDG. All blanks were free of any arsenic above the RL.
Cadmium SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for cadmium.
The cadmium analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7131A. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
The MS/MSD %Rs were within acceptance criteria except for as follows:
Sample AOC 48-SS03 (0-0.5�) | ||
Analyte | MSD %R | QC |
Cadmium | 74.3 | 80-122 |
The cadmium result in the associated samples from the same site with similar matrix as the MS/MSD sample was flagged �M� to indicate a matrix effect was present.
The LCS %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
All MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All initial and continuing calibration criteria were met. | |
All second source calibration criteria were met. | |
The five-fold diluted cadmium result was less than the reporting limit. Therefore, the dilution test was not applicable. | |
The recovery test criteria was not met for both soil and water: |
Sample AOC 48-SS03 (0-0.5�) | ||
Analyte | %R | QC |
Cadmium | 59.6 | 85-115 |
Sample RL83-EB09 | ||
Analyte | %R | QC |
Cadmium | 122 | 85-115 |
The cadmium result in the associated samples was considered estimated and flagged �J�.
There were two method blanks, one equipment blank and several calibration blanks associated with the cadmium analyses in this SDG. All blanks were free of any cadmium above the RL.
Lead SDG 32926
General
This SDG consisted of ten (10) samples, including six (6) confirmation environmental soil samples, one field duplicate soil sample, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on June 14, 2000 and analyzed for lead.
The lead analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7421. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.
Accuracy
Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG.
The MS/MSD %Rs were within acceptance criteria except as follows:
Sample AOC 48-SS03 (0-0.5�) | ||
Analyte | MS %R | QC |
Lead | -57.6 | 74-124 |
The lead result in the associated samples from the same site with similar matrix as the MS/MSD sample was flagged �M� to indicate a matrix effect was present.
The LCS %Rs were within acceptance criteria.
Precision
Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample AOC 48-SS03 (0-0.5�) was used as the MS/MSD sample in this SDG. Sample AOC 48-SS03 (0-0.5�) FD was the field duplicate of sample AOC 48-SS03 (0-0.5�).
The MS/MSD and field duplicate RPDs were within acceptance criteria.
Completeness
Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.
All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.
Representativeness
Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:
Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;
Comparing actual analytical procedures to those described in the AFCEE QAPP; | |
Evaluating holding times; and | |
Examining field and laboratory blanks for cross contamination of samples during sample collection and analysis. |
All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.
All initial and continuing calibration criteria were met. | |
All second source calibration criteria were met. | |
Since the five-fold diluted concentration exceeded the calibration range, the dilution test was done with 1/10 and 1/25 dilutions. All dilution test criteria were met. |
There were two method blanks, one equipment blank and several calibration blanks associated with the lead analyses in this SDG. The blanks were free of any lead above the RL.