[Home] ]

[Data Verification Reports]

RL33 Data Validation Summary Report

32337

Data Validators: Michelle Wolfe, Tammy Chang

Introduction

The following data validation summary report covers environmental soil samples and associated field quality control (QC) samples collected from the Camp Stanley CSSA Site B24 (under RL33) on March 30, 2000. Samples in the following laboratory Sample Delivery Group (SDG) were analyzed for semivolatile organic compounds (SVOCs); volatile organic compounds (VOCs); metals and mercury:

32337

 

 

Field quality control samples collected were trip blank; equipment blank; matrix spike/matrix spike duplicates (MS/MSD); and field duplicates. Parsons ES did not collect any ambient samples for VOC analysis due to the lack of determined source from the site. The trip blanks were analyzed for volatile organics only. All other field quality control samples were analyzed for the same parameters as their associated samples.

All samples were collected by Parsons Engineering Science (Parsons ES). All analyses were performed by APPL Laboratories, Inc. following procedures outlined in the AFCEE QAPP, version 3.0.

Evaluation Criteria

The data submitted by the laboratory has been reviewed and validated following the guidelines outlined in the AFCEE QAPP, version 3.0. Information reviewed in the data packages include sample results; the summary of laboratory quality control results; case narrative; raw data; and chain-of-custody forms. The analyses and findings presented in this report are based on the reviewed information, and whether guidelines in the AFCEE QAPP were met.

SVOC SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on March 30, 2000 and analyzed for semivolatile organic compounds (SVOCs).

SVOC analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 8270C. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples, LCS sample, and surrogate spikes. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD %Rs were within acceptance criteria except for the following:

Sample B24-SB05 (12.0-14.0�)

Analyte

MS %R

MSD %R

QC

Bis(2-ethylhexyl)phthalate

-676.5

-611.8

25-139

The bis(2-ethylhexyl)phthalate results in the samples from the same site and of similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.

All LCS and surrogate %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

All MS/MSD and field duplicate RPDs were within acceptance criteria except sample B24-SB05 (12.0-14.0�) contained 14 mg/kg of bis(2-ethylhexyl)phthalate and its field duplicate sample contained 0.26 mg/kg (less the RL) of the same compound. The %RPD calculation is not applicable.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

All results were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding times; and

Examining field and laboratory blanks for cross contamination of samples during collection or analysis.

All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All instrument performance check criteria were met.

All initial calibration criteria were met.

All continuing calibration criteria were met.

All second source verification criteria were met.

All internal standard criteria were met.

There were two method blanks and one equipment blank associated with the SVOC analyses in this SDG. All blanks were free of SVOCs above the RL.

VOC SDG 32337

General

This SDG consisted of fifteen (15) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples, one equipment blank sample and one trip blank sample. The samples were collected on March 30, 2000 and analyzed for volatile organic compounds (VOCs).

VOC analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 8260B. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding times required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples; LCS samples; and surrogate spikes. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD %Rs were within acceptance criteria except for the following:

Sample B24-SB05 (12.0-14.0�)

Analyte

MS %R

MSD %R

QC

1,2,3-trichlorobenzene

37.0

37.0

65-147

1,2,4-trichlorobenzene

38.9

37.0

65-145

1,2,4-trimethylbenzene

(68.5)

61.1

65-135

1,2-DCB

57.4

55.6

65-135

1,3-DCB

59.3

53.7

65-135

1,4-DCB

57.4

53.7

65-135

4-chlorotoluene

(66.7)

61.1

65-135

hexachlorobutadiene

53.7

33.3

65-135

n-butylbenzene

63.0

50.0

65-135

naphthalene

58.5

60.4

65-135

p-isopropyltoluene

(72.2)

57.4

65-135

sec-butylbenzene

(75.9)

59.3

65-135

tert-butylbenzene

(77.8)

63.0

65-135

 ( ) The %R was compliant.

The results for the non-compliant analytes in the samples from the same site and of similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.

All LCS and surrogate %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

All MS/MSD RPDs were within acceptance criteria except for the following:

Sample B24-SB05 (12.0-14.0�)

Analyte

RPD

QC

hexachlorobutadiene

46.8

30

The hexachlorobutadiene results in the samples from the same site and of similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.

All field duplicate RPDs were within acceptance criteria.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

All results were considered usable. The completeness for this SDG is 100.0% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding times; and

Examining field and laboratory blanks for cross contamination of samples during collection or analysis.

All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All instrument performance check criteria were met.

All initial calibration criteria were met.

All continuing calibration criteria were met.

All second source verification criteria were met.

All internal standard criteria were met.

There were two method blanks, one equipment blank and one trip blank associated with the VOC analyses in this SDG. The method blanks and trip blank were free of VOCs above the RL. The equipment blank, RL33-EB01, contained 5.90 mg/l of methylene chloride. The methylene chloride results in the samples collected the same day were flagged �B� to indicate a blank contamination was present.

Metals SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank. The samples were collected on March 30, 2000 and analyzed for metals; barium, chromium, copper, nickel, and zinc.

The barium, chromium, copper, nickel, and zinc analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 6010B. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD %Rs were within acceptance criteria except for the following:

Sample B24-SB05 (12.0-14.0�)

Analyte

MS %R

MSD %R

QC

nickel

74.0

69.0

75-125

The results for the non-compliant analytes in the samples from the same site and of similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.

All LCS %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

All MS/MSD and field duplicate RPDs were within acceptance criteria.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding times; and

Examining field and laboratory blanks for cross contamination of samples during collection and analysis.

All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All initial and continuing calibration criteria were met.

All second source calibration criteria were met.

All interference check criteria were met.

All dilution test criteria were met except for the following.

Sample B24-SB05 (12.0-14.0�)

Analyte

%RPD

QC

barium

16.0

10

The barium results in all samples in this SDG were considered estimated and flagged �J�. The chromium results were less than the reporting limit. The diluted results for copper, nickel, and zinc were less than the reporting limit. Therefore, the dilution test was not required for chromium, copper, nickel or zinc.

All post digestion spike addition criteria were met except for the following:

Sample B24-SB05 (12.0-14.0�)

Analyte

%R

QC

nickel

72.7

75-125

The nickel results in all samples in this SDG were considered estimated and flagged �J�.

There were two method blanks, one equipment blank and several calibration blanks associated with the metal analyses in this SDG. All blanks were free of any metals above the RL.

Mercury SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on March 30, 2000 and analyzed for mercury.

The mercury analyses were performed using USEPA SW846 Method 7470A for waters and 7471A for soils. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD and LCS %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

All MS/MSD and field duplicate RPDs were within acceptance criteria.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding ties; and

Examining field and laboratory blanks for cross contamination of samples during collection and analysis.

All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All initial and continuing calibration criteria were met.

All second source calibration criteria were met.

There were two method blanks, one equipment blank and several calibration blanks associated with the mercury analyses in this SDG. All blanks were free of any mercury above the RL.

Arsenic SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank. The samples were collected on March 30, 2000 and analyzed for arsenic.

The arsenic analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7060A. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD and LCS %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�)

The MS/MSD RPD was within acceptance criteria.

All field duplicate RPDs were within acceptance criteria except for the following:

B24-SB03 (0.0-0.5�) / FD

Analyte

Parent

FD

%RPD

QC

arsenic

3.60 mg/kg

1.59 mg/kg

77.5

15

The positive arsenic results in the samples collected on the same day as the field duplicate pair were considered estimated and flagged �J� and the non-detect results were considered unusable and flagged �R�.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

Six arsenic results were considered unusable and flagged �R� due to field duplicate outliers. The completeness for this SDG is 57.1% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP; Evaluating holding times; and

Examining field and laboratory blanks for cross contamination of samples during collection and analysis.

All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All initial and continuing calibration criteria were met.

All second source calibration criteria were met.

The diluted result for the sample used for the dilution test was less than the reporting limit. Therefore, the dilution test was not required.

The recovery test criteria was met.

There were two method blanks, one equipment blank and several calibration blanks associated with the arsenic analyses in this SDG. All blanks were free of any arsenic above the RL.

Cadmium SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on March 30, 2000 and analyzed for cadmium.

The cadmium analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7131A. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample B24-SB05 (12.0-14.0�) was used as the MS/MSD sample for this SDG.

All MS/MSD and LCS %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

The MS/MSD RPD was within acceptance criteria.

All field duplicate RPDs were within acceptance criteria except for the following:

B24-SB03 (0.0-0.5�) / FD

Analyte

Parent

FD

%RPD

QC

cadmium

0.28 mg/kg

0.23 mg/kg

19.6

15

The positive cadmium results in the samples collected on the same day as the field duplicate pair were considered estimated and flagged �J� and the non-detect results were considered unusable and flagged �R�.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

Six cadmium results were considered unusable and flagged �R� due to field duplicate outliers. The completeness for this SDG is 57.1% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions.

Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding times; and

Examining field and laboratory blanks for cross contamination samples during collection and analysis.

All samples in this SDG were analyzed following chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed with the holding times required for the analysis.

All initial and continuing calibration criteria were met.

All second source calibration criteria were met.

The diluted result for the sample chosen for the dilution test was less than the reporting limit. The dilution test was not required.

The recovery test failed to meet criteria as follows:

Sample B24-SB05 (12.0-14.0�)

Analyte

%R

QC

cadmium

115.7

85-115

The cadmium results in the associated samples were considered estimated and flagged �J�.

There were two method blanks, one equipment blank and several calibration blanks associated with the cadmium analyses in this SDG. All blanks were free of any cadmium above the RL.

Lead SDG 32337

General

This SDG consisted of fourteen (14) samples, including nine (9) confirmation environmental soil samples, two field duplicate soil samples, one set of matrix spike/matrix spike duplicate samples and one equipment blank sample. The samples were collected on March 30, 2000 and analyzed for lead.

The lead analyses were performed using United States Environmental Protection Agency (USEPA) SW846 Method 7421. All samples for this SDG were analyzed following the procedures outlined in the AFCEE QAPP. All samples collected were prepared and analyzed within the holding time required by the method.

Accuracy

Accuracy was evaluated using the %R results for the MS/MSD samples and LCS samples. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG.

All MS/MSD %Rs were within acceptance criteria except for the following:

Sample B24-SB04 (12.0-14.0�)

Analyte

MS %R

MSD %R

QC

lead

(84.8)

44.0

74-124

 ( ) %R was compliant

The lead results in the samples from the same site and of similar matrix as the MS/MSD sample were flagged �M� to indicate a matrix effect was present.

All LCS %Rs were within acceptance criteria.

Precision

Precision was evaluated using the Relative Percent Difference (RPD) results obtained from MS/MSD results; and the field duplicate analyte values. Sample B24-SB05 (12.0-14.0�) was analyzed as the MS/MSD sample for this SDG. Sample B24-SB03 (0.0-0.5�) FD was the field duplicate of sample B24-SB03 (0.0-0.5�). Sample B24-SB05 (12.0-14.0�) FD was the field duplicate of sample B24-SB05 (12.0-14.0�).

The MS/MSD RPD was within acceptance criteria.

All field duplicate RPDs were within acceptance criteria except for the following:

B24-SB05 (12.0-14.0�) / FD

Analyte

Parent

FD

%RPD

QC

lead

3.05 mg/kg

2.18 mg/kg

33.3

25

The positive lead result in the samples collected the same day as the field duplicate pair were considered estimated and flagged �J�.

Completeness

Completeness has been evaluated by comparing the total number of samples collected with the total number of samples with valid analytical data.

All results were considered usable. The completeness for this SDG is 100% compared to the minimum acceptance limit of 90%.

Representativeness

Representativeness expresses the degree to which sample data accurately and precisely represents actual site conditions. Representativeness has been evaluated by:

Comparing the chain-of-custody procedures to those described in the AFCEE QAPP;

Comparing actual analytical procedures to those described in the AFCEE QAPP;

Evaluating holding times; and

Examining field and laboratory blanks for cross contamination of samples during collection and analysis.

 All samples in this SDG were analyzed following the chain-of-custody forms (COCs) and analytical procedures described in the AFCEE QAPP. All samples were prepared and analyzed within the holding time required for the analysis.

All initial and continuing calibration criteria were met.

All initial and continuing calibration criteria were met.

All second source calibration criteria were met.

The dilution test failed to meet criteria as follows:

Sample B24-SB05 (12.0-14.0�)

Analyte

%D

QC

lead

16

10

The lead results in the associated samples were considered estimated and flagged �J�.

The recovery test failed to meet criteria as follows:

Sample B24-SB05 (12.0-14.0�)

Analyte

% R

QC

lead

132

75-125

The lead results in the associated samples were considered estimated and flagged �J�.

There were two method blanks, one equipment blank and several calibration blanks associated with the lead analyses in this SDG. All blanks were free of any lead above the RL.