[Home]  [Master Table of Contents]

[Meeting Minute Index]  [RL74 TIM #1 Minutes]

Attachment 4

Laboratory Status Conference Call Minutes

F11623-94-D0024/RL17, RL33, and RL53

F41624-94-D-8136/DO23

Date:

March 11, 1999

Time:

10:30 A.M. - 12:00 P.M.

Place:

Parsons Engineering Science, Austin office

Subjects:

Laboratory Status

Participants:

Participant

Organization

Brian K. Murphy

CSSA ENV

Jo Jean Mullen

AFCEE/ERD

Rene G. Hefner

AFCEE/ERC

Nancy Stine

AMC CONF

Dave McCollough

AMC CONF

Ma Thong

AMC CONF

Beth Shively

AFCEE/ERC

Garry Earls

Parsons ES, St. Louis

Greg Jones

Parsons ES, Atlanta

Peter Holland

Parsons ES, Austin

Julie Burdey

Parsons ES, Austin

Susan Roberts

Parsons ES, Austin

Karuna Mirchandani

Parsons ES, Austin

Tammy Chang

Parsons ES, Austin

Minutes prepared by Julie Burdey, Parsons ES.

Prior to conferencing in participants from the contracting office of Air Mobility Command (AMC CONF), Parsons ES-St. Louis, and Parsons ES-Atlanta, there was some discussion regarding the U.S. Environmental Protection Agency’s (EPA’s) investigation of ITS Laboratory.  This discussion is summarized in the following paragraphs:

Pre-Conference Discussion

Brian Murphy, CSSA, indicated that he recently spoke with Guy Tidmore at EPA regarding their investigation of ITS.  Mr. Tidmore told Brian that at least four, and up to eight, people may be indicted at ITS.  Mr. Tidmore also told Brian that EPA plans to not accept any VOC or SVOC data from ITS, but they haven’t found a problem with metals data from ITS.  Jo Jean Mullen, AFCEE/ERD, said that, during AFCEE’s audit of ITS, ITS personnel reported that they weren’t following their standard operating procedures (SOP).  However, Mr. Murphy responded, we cannot prove that they weren’t.  He added that legally, we can’t not accept the metals data.  Ms. Mullen responded that, if ITS personnel say in court that they didn’t follow the SOP or the project required procedures, then metals data can be thrown out. 

Mr. Murphy said that the U.S. Army Corps of Engineers (USACE) is going to have to spend $200 million to fix their problem associated with the fraudulent ITS data.  He added that most contractors don’t know the extent of the problem.  Ms. Mullen said that Earth Tech and Rust are now going through what Parsons ES did.  When they request information from ITS, ITS only sends part of it, then they have to ask for it again, etc.  Ms. Mullen said that Earth Tech’s lawyers are going to say that ITS is in breach of contract.

Mr. Murphy added that he mentioned to Mr. Tidmore at EPA that ITS was trying to get CSSA, AFCEE, and Parsons ES to change data quality objectives (DQOs).  Mr. Tidmore said that EPA is going to come out with a position on ITS.  The position letter will be sent to ITS, but Mr. Tidmore will send a copy to Mr. Murphy, who will send a copy to AFCEE and Parsons ES.  Mr. Murphy also said that Mr. Tidmore told him that EPA has audited other labs to make sure they are not following the same fraudulent practices that ITS was, and they haven’t found any labs doing that.

Conference Call

The conference call was initiated with Ms. Mullen’s description of her concerns regarding the February 26 letter from Garry Earls, Parsons ES Program Manager, to Nancy Stine, AMC Contracting Officer.  The letter addressed Parsons ES’ proposal to resolve the ITS data problem at CSSA.  Ms. Mullen has concerns regarding the second paragraph on page 2 of the letter which implies that AFCEE’s QAPP requirements are too stringent.  Ms. Mullen feels that this statement is inappropriate and that the Parsons ES Program Manager does not have a technical understanding as to why the DQOs are set as they are.

Ms. Mullen added that she would like to allow laboratories to follow EPA contract laboratory program (CLP) requirements, but CSSA can’t use CLP data.  Ms. Stine then stated that Parsons is having difficulty finding a laboratory that can meet the AFCEE QAPP requirements.  Ms. Mullen responded that she did not think that is necessarily accurate.  She said she has not seen all of the variances from all of the laboratories.  AFCEE will consider variances, as long as they aren’t outlandish.  She added that final approval of variances would ultimately have to come from the regulatory agencies.  After EPA comes on board, AFCEE won’t be able to allow any variances without approval from EPA.

At this point, the group agreed to discuss one laboratory (O’Brien & Gere) that is being considered for work under DO23 of the AFCEE contract.  Susan Roberts noted that, although this laboratory was not currently being considered for work under orders RL33 and RL53 of the AMC contract, this discussion would serve as a “case study” for the types of problems Parsons ES is experiencing regarding laboratory subcontracting.  Requests for proposal (RFPs) for RL33 and RL53 were sent to four additional laboratories on March 4 based on input from AFCEE.  Bids and variances were received on March 10, which did not allow sufficient time for Parsons chemists to review the packages prior to this conference call.

Regarding the O’Brien & Gere package, Ms. Mullen said that the preliminary data package that the laboratory sent was basically a CLP package.  O’Brien & Gere’s variances said that they would use CLP-like forms, and also included text which implied use of CLP procedures.

Ms. Roberts then introduced tables that Parsons ES prepared to summarize all of the major variances received by labs proposing on the CSSA analytical work.  These tables are provided as Attachment A.  Karuna Mirchandani indicated that the O’Brien & Gere package submitted to AFCEE was CLP-like, but that the format is still under discussion.  O’Brien & Gere is currently trying to work through AFCEE’s concerns regarding the format.  She added that technical issues associated with O’Brien & Gere include elevated reporting limits (RLs) and wider quality control (QC) acceptance criteria.

Ms. Mullen responded that those analytes for which O’Brien & Gere have requested elevated RLs, are the same analytes for which they want wider QC acceptance criteria.  Ms. Mullen said that they can have one or the other.  She added that, if you look at the CLP program and compare the reporting limits and acceptance criteria allowed by that program, you’ll notice the numbers that O’Brien & Gere has requested in their variances are the same as the CLP numbers, assuming they are even following CLP.

Ms. Mirchandani indicated that Parsons could request O’Brien & Gere’s SOPs and control charts for each of the parameters we are reviewing.  Parsons could review each of the SOPs to make sure that O’Brien & Gere is not just following CLP requirements, and that they have a history of needing the wider QC acceptance criteria.  Ms. Mullen responded that this effort would take a lot of time for Parsons, and that it would be difficult to schedule around this.  Ms. Mirchandani clarified that Parsons would only review each compound for which a variance has been requested.  Ms. Roberts added that Parsons wouldn’t know how long this would take until Parsons received the SOPs.

Greg Jones, Parsons ES-Atlanta, added that Parsons ES needs to determine if O’Brien & Gere is just providing a CLP-like package, or if they are following CLP procedures.  He added that CLP procedures have side limits designed for pushing through a lot of data of known quality, not necessarily of high quality.  Parsons would have to verify that O’Brien & Gere’s SOPs follow SW-846.  Mr. Jones said that lab’s SOPs are usually based on SW-846, but they can also be based on CLP requirements.  O’Brien & Gere has told Parsons that they are following SW-846 but Parsons has not reviewed their SOPs to verify this.

Ms. Mullen added that SW-846 provides guidelines.  Laboratories say that they are using SW-846, but in reality they are mixing SW-846 and CLP, and taking the loosest requirements from each.  Some examples include using the practical quantitation limit (PQL) as the nondetect (ND) value; not using the AFCEE “U” or “F” flag; applying the “U” flag where AFCEE would have none; and adjusting method detection limits (MDLs) and reporting limits (RLs) for dry weight.  Ms. Mullen added that these are very basic differences that go to the heart of reporting, and that AFCEE does not allow CLP reporting, period.

Mr. Earls noted that it sounded like O’Brien & Gere is not a viable laboratory for the CSSA project.  Ms. Mullen responded that she didn’t know.  O’Brien & Gere is in Dames & Moore’s contract laboratory program and they say O’Brien & Gere can do AFCEE QAPP analytical testing.  O’Brien & Gere has said themselves that they can do it.  However, by looking at the variances, Ms. Mullen said she can’t tell.

Mr. Jones indicated that, from the standpoint of a laboratory manager, if the laboratory has requested variances, that means that if they don’t get the variances approved, they can’t do it.  Ms. Mullen responded that that isn’t necessarily true.  AFCEE has seen other labs revise their variance requests based on specific project needs.  Mr. McCollough then asked if O’Brien & Gere might be acceptable based on a more rigorous demonstration of their need for these variances.

Ms. Mullen responded that some variances would not be accepted by AFCEE.  AFCEE will not accept CLP reporting, CLP flagging, nor CLP QA/QC.  At this point, all agreed to discuss each variance requested by O’Brien & Gere, summarized in Tables 2, 3, and 4 in Attachment
4-A
.  The results of these discussions regarding specific O’Brien & Gere variances are summarized in Attachment 4-B.

After each of the variances were discussed as noted in Attachment 4-B, Ms. Roberts said that Parsons ES would move forward to try to get O’Brien & Gere on board for the CSSA project.  Ms. Mullen added that she felt that Parsons could handle whatever O’Brien & Gere can’t do, and that CSSA is willing to pay the additional costs.  Parsons just needs to notify AFCEE of the additional effort involved. 

Ms. Stine then asked what the Parsons plan is regarding the rework for RL17 and RL33, and when AMC can expect a schedule for the rework.  Ms. Roberts responded that Parsons can put together a general schedule, but it is difficult to estimate how long actions will take.  Our prior experience cannot be used to estimate how long it will take to get these laboratory issues resolved.  Parsons can provide a schedule which includes assumptions about how long it will take for each step, but prefers not to give specific dates, if that is acceptable to AMC and AFCEE.  Ms. Mullen agreed that would be adequate for CSSA’s needs. 

Ms. Stine also asked if AFCEE, CSSA, and Parsons all agree on the level of effort required for the resampling (i.e. is Parsons going to resample everything?).  Ms. Mullen responded that complete resampling may not be necessary.  She indicated that input from the regulators would be required.  Ms. Roberts indicated that Parsons would likely propose no resampling at locations which had high levels of contaminants detected, since ITS’ problem affects non-detects and low level detects.  Ms. Mullen agreed, and reiterated that approval for the reduced resampling effort must be obtained from the regulatory agencies.

Ms. Roberts indicated that the schedule for the ressampling effort will be provided on March 31, 1999, and AFCEE and AMC agreed to this date. 

Mr. Murphy indicated that CSSA would like to get two labs on board for soils, for a total of three laboratories.  Mr. Earls noted that there would be additional expense involved with that.  Mr. Murphy agreed, and asked that Parsons ES submit a cost estimate for additional lab procurement along with the schedule.

Ms. Stine asked how the rework is being charged.  Mr. Earls responded that Parsons has set up its own internal job number for these efforts.  Parsons is trying to pursue ITS, ITS’ insurance company, and ITS’ parent company for compensation. 

Mr. Murphy repeated some of the information that was discussed during the pre-conference discussion.  He noted that EPA will be indicting 4 to 8 people at ITS; that EPA has found the VOC and SVOC data to be invalid; and that EPA has not found anything wrong with the metals data.  Mr. Earls noted that Parsons would like to use the maximum amount of ITS data that meets CSSA’s requirements.

Attachment 4-A

Table 1.  List of Labs Responding to RFP for DO23

Project

Laboratory Name and Date Submitted

General Comments

Proposed Cost

Report Forms

Format

Electronic Deliverable

Flag

DO23

Savannah (Feb 98)

Declined to bid.

No Bid

 

 

 

DO23

Ceimic (Feb 98)

Has never done AFCEE work according, to Ceimic contact.

$50,770

CLP - like

 

 

DO23

DHL Analytical (Feb 98)

Declined to bid.

No Bid

 

 

 

DO23

Ecology (Sept 98)

Can’t provide results for full AFCEE list of compounds.

$25,840

Reporting unit for soil: µg/kg, for water: µg/L

 

 

DO23

EMAX (Sept 98)

Can’t do AFCEE flags, did not meet variances requested on another AFCEE project.

$34,036

In-house reporting formats

 

Different flags

DO23

O’Brien & Gere (Sept 98)

In Dames & Moore contract lab program (CLP), currently doing AFCEE work, cooperative in providing information. Consider to use.

$33,474

All positive results with 2 sig. fig.  In-house report formats

Does not include TICs

  

DO23

Parsons ES Lab (Feb 98)

Declined to bid.

No Bid

 

 

 

Note:  Shaded line represents laboratory preferred by Parsons ES for project analyses.

Attachment 4-A (continued)

Table 1.  List of Labs Responding to RFP for RL33 and RL53

Project

Laboratory Name and Date Submitted
General Comments

Proposed Cost

Report Forms

Format

Electronic Deliverable

Flag

RL33/RL53

Accutest (Dec 98)

Following SW-846 instead of AFCEE QAPP 3.0. No attempt to meet AFCEE QAPP requirements.

$49,195

CLP - like

 

CLP - flag

RL33/RL53

APPL (Mar 99)

Consider for use based on latest preliminary response to RFP received March 10, 1999.

NA

 

 

 

RL33/RL53

Chemron (July 98)

Missed holding times and misreported results to CSSA.

$44,760

 

 

 

RL33/RL53 

Columbia  (Dec 98)

Declined to bid.

No Bid

 

 

 

RL33/RL53

Data Chem Lab (Feb 98, Mar 99) 

Declined to bid on first round.  Consider for use based on latest preliminary response to RFP received March 10, 1999.

$56,230

 

  

  

RL33/RL53

EA (Dec 98)

Declined to report AFACEE format unless variances were granted, poor customer service, no historical control charts.

$41,304 + 20% (to manage additional AFCEE QA/QC issues)

CLP–like Currently the lab is working on the development of AFCEE QAPP 3.0 report formats.

  

  

RL33/RL53

Kemron (Mar 99)

Consider for use based on latest preliminary response to RFP received March 10, 1999.

NA

 

 

 

RL33/RL53

Paragon (Dec 98) 

Poor reference. 

$45,058

Level IV CLP - like

 

 

RL33/RL53

PDP (Mar 99)

Submitted inadequate variances.

$50,230

PDP does not have QAPP 3.0 forms. Currently under development. 

 

 

RL33/RL53

QA (July 98)

No longer exists.

$40,383

 

 

  

RL33/RL53

Quanterra (July 98)

Missed holding times and turn-around times, poor customer service on other Parsons projects.

$43,623

AFCEE QAPP 2.0

 

 

Note:  Shaded line represents laboratory preferred by Parsons ES for project analyses.  Since RFPs were recently sent to labs and responses are not completely in, final determination of recommended laboratory is not yet possible.

Attachment 4-A (continued)

Table 2.  Summary of Variances for Metals and Explosives

 

 

SW-7000 Metals

SW7470A/7471A

SW-8330

 

Reporting Limits

Lab Control Sample

Calibration

Reporting Limits

Substitute with ICP method

Reporting Limits

Post-spike & Method of Standard Addition (MSA)

Reporting Limits

Surrogates

Calibrations

Lab Control Sample

Confirmation

Accutest

Follows SW-846, use the lowest calibration concentration as the RL

Only use to determine method compliance if MS/MSD fails

Use one standard and one blank

Higher RLs

 

 

No MSA, no matrix check dilutions

Based on SW 846

 

 

 

 

APPL

Zn: 0.05 ppm for water and 5.0 ppm for soil

 

 

 

Cd

 

 

 

 

 

 

 

Ceimic

W: 7 elements with higher RL

 

 

 

 

 

 

 

 

 

 

 

Chemron

 

Do not reanalyze sample when LCS is high but sample shows ND

 

No MDL study for Cd

 

 

 

 

 

 

 

 

Columbia  (declined)

--

--

--

--

--

--

--

--

--

--

--

--

Data Chem

 

 

 

 

 

 

 

Higher RL for 3 nitrotoluene isomers, reports o- and p-nitrotoluene together

 

 

Allow one of 7 compds with 75% as LCL to be lowered to 50%

 

DHL Analytical  (declined)

--

--

--

--

--

--

--

--

--

--

--

--

EA

RL verification: %R 50%-150%

Different corrective action

 

W: 2 compds with higher RL

S: 1 compd with higher RL

RL and LCS have same variances as 6010B 

 

W: higher RL

 

 

Only reanalyze when LCS, MB and sample failed

Follow SW8000B Sec. 7.7

80% of analytes meet acceptance criteria

2nd column confirmation does not have to meet the acceptance criteria

Ecology

S: Zn is missing from the list

 

 

 

Pb, As, Cd

 

 

 

 

 

 

 

EMAX

W: 6 elements with higher RL S: 3 elements with higher RL

 

 

 

 

 

 

 

 

 

 

 

Kemron (bid not yet available)

 

 

 

 

 

 

 

 

 

 

 

 

Attachment 4-A (continued)

Table 2.  Summary of Variances for Metals and Explosives

 

 

SW-7000 Metals

SW7470A/7471A

SW-8330

 

Reporting Limits

Lab Control Sample

Calibration

Reporting Limits

Substitute with ICP method

Reporting Limits

Post-spike & Method of Standard Addition (MSA)

Reporting Limits

Surrogates

Calibrations

Lab Control Sample

Confirmation

O’Brien & Gere

1:  Zn has higher RL than AFCEE

 

 

 

 

2:  MDL> half of AFCEE’s RL

 

 

 

 

 

 

Paragon

W: 3 elements with higher RL

S: 2 elements with higher RL

 

 

 

 

 

 

 

 

 

 

 

Parsons ES Lab (declined)

--

--

--

--

--

--

--

--

--

--

--

--

PDP

 

 

 

 

 

 

 

 

 

 

 

 

QA (closed)

 

 

 

 

 

 

 

 

 

 

 

 

Quanterra

 

 

 

 

 

 

 

 

 

 

 

 

Savannah (declined)

--

--

--

--

--

 

 

--

--

--

--

--

Attachment 4-A (continued)

Table 3.  Summary of Variances for Volatile Organic Compounds

 

SW 8260B

Reporting Limits

Laboratory Control Sample

Initial Calibration Verification

Continuing Calibration Verification

Internal Standards

2nd source Calibration Verification

Surrogates

Method Blank

Retention Time

System Performance Check Compounds

Instrument Tuning

Accuracy

Accutest

Follows SW-846, use the lowest calibration concentration as the RL

Only use to determine method compliance if MS/MSD fails

 

Different acceptance criteria

 

 

Do not automatically re-extract when surrogate fails

 

Follow criteria for RT in the SW 846

 

 

 

APPL

W: 20 compds with higher RL

S: 4 compds with higher RL

 

 

 

 

 

 

 

 

 

 

 

Ceimic

1 ppb for water; 5 ppb for soil

Only reanalyze sample when %R is higher than the UCL and analyte is found in the sample

If %RSD>15%, use cc of linear regression  or avg. RF  to check

1 ppb as the lowest cal std and 2 ppb for 1,2-dibromo-3-chloropropane (w); no corrective action if %Drift<30% and samples do not contain target analytes

 

 

 

Increase limit for MeCl2 to 5 ppb

 

 

 

Many variances based on historically developed control limits

Chemron

Report m- & p-Xylene together

Only reanalyze when sample > RL

 

 

 

 

 

Only reanalyze when MB > RL

 

 

 

 

Columbia (declined)

--

--

--

--

--

--

--

--

--

--

--

--

Attachment 4-A (continued)

Table 3.  Summary of Variances for Volatile Organic Compounds

 

SW 8260B

 

Reporting Limits

Laboratory Control Sample

Initial Calibration Verification

Continuing Calibration Verification

Internal Standards

2nd source Calibration Verification

Surrogates

Method Blank

Retention Time

System Performance Check Compounds

Instrument Tuning

Accuracy

DHL Analytical (declined)

--

--

--

--

--

--

--

--

--

--

--

--

Data Chem

 

 

0.5 ppb is the lowest calibration point, includes a 0.3 ppb RL check

0.5 ppb is the lowest calibration point; includes a 0.3 ppb RL check; uses mean RF instead of mean RSD; allows up to 4 analytes >20% but <30%

 

 

Uses 3 surrogates

Higher RL for MeCl2

 

 

 

 

EA

W: 29 compds with higher RL; S: 10 compds with higher RL

80% of analytes meet acceptance criteria

 

Follow SW8000B Sec 7.7

Wider EICP area

Follow SW8000B Sec 7.7

Only reanalyze when MB, LCS and sample are failed

Wider range for MeCl2 and acetone

 

Dec. 1 mini. RF requirement;  Follow SW8000B Sec. 7.5

Alternative tuning criteria

 

Ecology

Report m-, p-Xylene together;  Higher RL for 3 compds; 15 compds are missing from the list

modified MDL procedure

No reanalyze when LCS is high

 

Lowest cal pt. for water is 1 ppb and for soil is 5 ppb except ketones

 

 

 

Higher limits for acetone and MeCl2

 

 

 

 

Attachment 4-A (continued)

Table 3.  Summary of Variances for Volatile Organic Compounds

 

SW 8260B

 

Reporting Limits

Laboratory Control Sample

Initial Calibration Verification

Continuing Calibration Verification

Internal Standards

2nd source Calibration Verification

Surrogates

Method Blank

Retention Time

System Performance Check Compounds

Instrument Tuning

Accuracy

EMAX

Report m- and p-Xylene together; W: 14 compds with higher RL

 

 

Only apply to analytes of concern

Uses 4 IS instead of 3

Only apply to analytes of concern

Uses 3 surrogates instead of 4

 

 

 

 

 

Kemron (bid not yet evaluated)

--

--

--

--

--

--

--

--

--

--

--

--

O’Brien & Gere

3a:  W: 4 compds with higher RL

3b:  S: 4 compds with higher RL

3C:  report m-, p-Xylene together

3d:  adjusted by dry wt.

 

 

 

 

4:  50-150%R for dichlorodi-fluoromethane

5:  No reextraction for med-level soil samples with one surrogate failed

6:  Inc limit for MeCl2 to 1 ppb

 

 

 

7:  50 - 150%R for dichloro-difluoro-methane

Paragon

Higher limit for MeCl3

Wider range for MeCl3

%RSD<50% for MeCl3

MeCl3 +/- 30%

Use different IS

MeCl3 +/- 30%

 

 

 

 

 

 

Parsons ES Lab (declined)

--

--

--

--

--

--

--

--

--

--

--

--

PDP

Uses the lowest calibration point as the RL; Report m-&p-Xylene together

 

 

 

 

 

 

 

 

 

 

 

Attachment 4-A (continued)

Table 3.  Summary of Variances for Volatile Organic Compounds

 

SW 8260B

 

Reporting Limits

Laboratory Control Sample

Initial Calibration Verification

Continuing Calibration Verification

Internal Standards

2nd source Calibration Verification

Surrogates

Method Blank

Retention Time

System Performance Check Compounds

Instrument Tuning

Accuracy

QA

 

 

 

 

 

 

 

 

 

 

 

 

Quanterra

Report m- & p-Xylene together; W: 19 compds with higher RL; S: 2 compds with higher RL

 

 

 

 

 

 

 

 

 

 

 

Savannah (declined)

--

--

--

--

--

--

--

--

--

--

--

--

Attachment 4-A (continued)

Table 4.  Summary of Variances for Semivolatile Organic Compounds

Laboratories

SW-8270C

Reporting Limits

Initial Calibration Verification

Continuing Calibration Verification

Surrogates

Retention Time

2nd Source Calibration Verification

Internal Standards

Method Blank

Laboratory Control Sample

Instrument Tuning

Accuracy

Accutest

Follows SW-846, use the lowest calibration concentration as the RL

 

20%D of ICV

 

Follow the RT criteria in SW 846

 

 

 

Only be used to determine method compliance if MS/MSD failed

 

 

APPL

 

 

 

 

 

 

 

 

 

 

 

Ceimic

 

 

 

 

 

 

 

 

 

 

 

Chemron

Report 3- & 4-Methyl phenol  together

 

 

 

 

 

 

 

Do not reanalyze sample when LCS is high but sample shows ND

 

 

Columbia (declined)

--

--

--

--

--

--

--

--

--

--

--

DHL Analytical (declined)

--

--

--

--

--

--

--

--

--

--

--

Data Chem

 

 

 

 

 

 

 

 

 

 

 

EA

 

Follow SW8000B Sec. 7.7

Follow SW8000B Sec. 7.7

Only reanalyze when MB, LCS  and sample failed

 

Follow SW8000B Sec. 7.7

Wider EICP acceptance range

exception for all common phthalates

80% of analytes meet acceptance criteria

Alternative tuning

W: 5 compounds

S: 5 compounds

Ecology

3 compounds with higher RL

 

 

 

 

 

 

 

 

 

 

Attachment 4-A (continued)

Table 4.  Summary of Variances for Semivolatile Organic Compounds

 

SW-8270C

Laboratories

Reporting Limits

Initial Calibration Verification

Continuing Calibration Verification

Surrogates

Retention Time

2nd Source Calibration Verification

Internal Standards

Method Blank

Laboratory Control Sample

Instrument Tuning

Accuracy

EMAX

 

Only apply to analytes of concern

Only apply to analytes of concern

 

 

Only apply to analytes of concern

 

 

Only apply to analytes of concern

 

 

Kemron (bid not yet available)

 

 

 

 

 

 

 

 

 

 

 

O’Brien & Gere

8:  Adjusted by dry wt.

 

 

 

 

 

 

 

 

 

 

Paragon

W: 2 compounds with higher RL

 

30%D of ICV for 6 compounds

 

Eliminated

 

Compare to CCV instead of ICAL

 

 

 

Wider ranges for 6 compounds (both w and s)

Parsons ES Lab (declined)

--

--

--

--

--

--

--

--

--

--

--

PDP

 

 

 

 

 

 

 

 

 

 

 

QA (closed)

--

--

--

--

--

--

--

--

--

--

--

Quanterra

 

 

Higher %R for  11 poor performers

 

 

+/-30% for 11 poor performers

 

 

 

 

Wider range for hexachlorocyclo-pentadiene
(10-125%)

Savannah (declined)

--

--

--

--

--

--

--

--

--

--

--

Attachment 4-B

O’Brien & Gere Variances and Conference Call Agreements

Cross-Reference No. from Att. A

Section and Page in AFCEE QAPP 3.0

Variance

O’Brien & Gere Justification

Conference Call Discussion

Final Decision from AFCEE

Detection Limits:

2

4.3.2 Reporting Limits, page 4-6

The following MDL is more than one-half of the RL:  mercury (soil)

The MDL (0.06 mg/kg) does not support the requirement of one half the RL (0.1 mg/kg).  However, the AFCEE RL can still be met.

This comment addresses soil reporting limits, but the contract currently planned under DO23 is only for analysis of groundwater samples; therefore, this variance does not apply.  Furthermore, there are no established background levels for groundwater at CSSA.  For soil analyses, the established background “PQL” is 0.1 mg/kg.  O’Brien and Gere’s required MDL of 0.06 mg/kg is only slightly above the AFCEE requirement of 0.05 mg/kg.

Approved.

1

7.2.15-1 Reporting Limits for Method SW6010B, page 7‑106

Lab can meet the required soil RL’s, with the following exception: Zinc   2.0 mg/kg

Based on the MDL study submitted, RLs must be elevated slightly to achieve AFCEE’s requirement of MDLs one half of the RLs.

The RL listed in AFCEE QAPP 3.0 for zinc (method 6010B) is 1.0 mg/kg.  However, the background “PQL” value is 2.0 mg/kg, the value that O’Brien & Gere is proposing.  Although this does not apply to the analyses planned under DO23, this variance is acceptable to AFCEE.

Approved.

3a

7.2.9-1 Reporting Limits for Method SW8260B, page 7‑59, 60

Lab can meet the required aqueous RL’s, with the following exceptions: 1,1,2,2-Tetrachloroethane   0.5 ug/l

Methylene chloride   2.0 ug/l

Naphthalene  1.0 ug/l

Styrene   0.5 ug/l

1,1,2,2-Tetrachloroethane:  A reactive compound and the low standards can have poor response, affecting the linearity.

Methylene chloride:  A common laboratory contaminant.  If the RL is too low, laboratory background may elevate the low standard response and result in poor linearity.

Naphthalene:  Moderate concentrations in analytical samples and standards are prone to cross-contamination in ensuing samples.  Higher RL reduces the possibility of a false positive being reported.
Styrene:  A reactive compound and the low standards can have poor response, affecting the linearity.

AFCEE RLs and TNRCC risk reduction standards  for the four compounds are as follows: 1,1,2,2-Tetrachloroethane:  RL-0.4 ug/l, RRS-4.3 ug/l

Methylene chloride:  RL-0.3 ug/l, RRS-5 ug/l

Naphthalene: RL-0.4 ug/l, RRS-730 ug/l

Styrene:  RL-0.4 ug/l, RRS-100 ug/l

Ms. Mullen indicated that, since methylene chloride, naphthalene, and styrene are not contaminants of concern, the elevated RLs for these compounds would be acceptable.  The requested RL for 1,1,2,2-tetrachloroethane is only 0.1 ug/l above the AFCEE RL; therefore, it is also acceptable.  Furthermore, all of the requested RLs are below the TNRCC risk reduction standards.

Approved.

Attachment 4-B

O’Brien & Gere Variances and Conference Call Agreements

Cross-Reference No. from Att. A

Section and Page in AFCEE QAPP 3.0

Variance

O’Brien & Gere Justification

Conference Call Discussion

Final Decision from AFCEE

Detection Limits:

3b

7.2.9-1 Reporting Limits for Method SW8260B, page 7‑59, 60

Lab can meet the required soil RL’s, with the following exceptions:

1,1,2,2-Tetrachloroethane  0.0025 mg/kg

Methylene chloride   0.005 mg/kg Naphthalene  0.005 mg/kg

Styrene   0.0025 mg/kg

1,1,2,2-Tetrachloroethane:  A reactive compound and the low standards can have poor response, affecting the linearity.

Methylene chloride:  A common laboratory contaminant.  If the RL is too low, laboratory background may elevate the low standard response and result in poor linearity.

Naphthalene:  Moderate concentrations in analytical samples and standards are prone to cross-contamination in ensuing samples.  Higher RL reduces the possibility of a false positive being reported.

Styrene:  A reactive compound and the low standards can have poor response, affecting the linearity.

AFCEE RLs and the TNRCC risk reduction standards (most conservative value from GWP and SAI res. and ind. listed) for the four compounds are as follows:

1,1,2,2-Tetrachloroethane:  RL-0.002 mg/kg, RRS-0.43 mg/kg

Methylene chloride:  RL-0.002 mg/kg, RRS-0.5 mg/kg

Naphthalene:  RL-0.002 mg/kg, RRS-73 mg/kg

Styrene: RL- 0.002 mg/kg, RRS-10 mg/kg

Ms. Mullen indicated that, since methylene chloride, naphthalene, and styrene are not contaminants of concern, the elevated RLs for these compounds would be acceptable.  The requested RL for 1,1,2,2-tetrachloroethane is only 0.1 ug/l above the AFCEE RL; therefore, it is also acceptable. Furthermore, all of the requested RLs are below the TNRCC risk reduction standards.

Approved.

Method QC Procedures:  Method 8260B

3c

Table 7.2.9-1.  RLs for Method SW8260B, page 7-60

m and p-Xylene will be reported together as (m+p)-Xylene with an RL of:
waters  0.6 ug/l
soils  0.004 mg/kg

m and p-Xylene cannot be resolved on capillary columns commonly utilized for SW8260 methods.

Approved without discussion.

Approved.

7

Table 7.2.9-2.  QC Acceptance Criteria for Method SW8260B, page 7-61

Use water and soil control limits of 50-150 (%R) for accuracy for dichlorodifluoromethane.

This analyte has been historically very difficult to purchase from two separate vendors and meet criteria of 75-125 %R.

AFCEE’s required control limits are
65-135 for soils
75-125 for water
Parsons noted that 50-150% is not unusual for a lab because dichlorodifluoromethane (freon) is a gaseous compound of which trace amounts can be found in air.  Ms. Shively, AFCEE, also noted that it probably is not an analyte of concern at CSSA.

Approved.
Parsons will request O’Brien and Gere to provide control charts for this analyte.

Attachment 4-B

O’Brien & Gere Variances and Conference Call Agreements

Cross-Reference No. from Att. A

Section and Page in AFCEE QAPP 3.0

Variance

O’Brien & Gere Justification

Conference Call Discussion

Final Decision from AFCEE

Method QC Procedures: Method 8260B

4

Table 7.2.9-3.  Summary of Calibration and QC Procedures for Method SW8260B:  Second source calibration verification page 7‑63

Use water and soil control limits of 50-150 (%R) for dichlorodifluoromethane.

This analyte has been historically very difficult to purchase from two separate vendors and meet criteria of 75-125 %R.

AFCEE’s required control limits are 65-135 for soils 75-125 for water Parsons noted that 50-150% is not unusual for a lab because dichlorodifluoromethane (freon) is a gaseous compound of which trace amounts can be found in air.  Ms. Shively, AFCEE, also noted that it probably is not an analyte of concern at CSSA.

Approved.

Parsons will request O’Brien and Gere to provide control charts for this analyte.

6

Table 7.2.9-3.  Summary of Calibration and QC Procedures for Method SW8260B:  method blank page 7‑65

Increase the method blank acceptance criteria of methylene chloride to 1.0 ppb.

This is a common laboratory contaminant.

Approved without discussion.

Approved.

5

Table 7.2.9-3.  Summary of Calibration and QC Procedures for Method SW8260B:  Surrogate Spike page 7‑67

Medium-level soil samples will not be re-extracted for the failure of one surrogate.  The sample extract will be re-injected only.

The failure of one surrogate is an indication of matrix effects.  The sample extract will be re-injected and flagged according to the AFCEE flags.

Ms. Shively at AFCEE noted that this variance is a request to allow O’Brien & Gere to do something differently than what other labs are allowed to do.  Mr. Jones noted that really only the analyst sitting at the bench can tell if there is a matrix effect.  CONDITION:  Ms. Shively indicated that the surrogate should be flagged as required by the QAPP, but that an explanation be given in the case narrative.

Approved with condition.  Follow AFCEE QAPP requirements then note in case narrative.  Parsons ES to request that O’Brien & Gere clarify their variance.

Reporting Requirements:

3d, 8

8.2  Data Review, Validation, and Reporting Requirements for Definitive Data, p. 8‑2

The AFCEE QAPP requires that MDLs and results be reported to one decimal place more than the corresponding RL.  O’Brien & Gere request a variance to report all positive results to two significant figures.

All of our LIMS data is formatted in this manner.  Results are reported to 2 significant figures or the RL.  The results are not significant past the RL.  (e.g., RL=10, result reported as 11 not 11.1).  MDLs will be reported electronically to the calculated MDL as defined in section 4, 4.3.1 (p. 4-5).

Not discussed during conference call.

None.

Attachment 4-B

O’Brien & Gere Variances and Conference Call Agreements

Cross-Reference No. from Att. A

Section and Page in AFCEE QAPP 3.0

Variance

O’Brien & Gere Justification

Conference Call Discussion

Final Decision from AFCEE

Reporting Requirements

 

8.2  Data Review, Validation, and Reporting Requirements for Definitive Data, p. 8‑2

The AFCEE QAPP requires that RLs and MDLs not be adjusted for dry weight.  O’Brien & Gere request a variance to report MDLs and RLs which have been adjusted for dry weight.  A wet weight aliquot of sample, as specified by the method, will be used for analysis then dry weight converted based on the percent solids.  RLs and MDLs are then dry weight converted.  RLs and MDLs are provided electonically only.

The industry stsandard is to take a wet weight aliquot of soil sample.

Ms. Mullen indicated that AFCEE is not having a problem with the numbers, but with the adjustment.  Ms. Mirchandani noted that labs typically adjust MDLs and RLs.  Labs’ LIMS systems are set up to flag prior to correction, then to report the adjusted limits on the reporting form.  Ms. Roberts asked that if O’Brien & Gere can’t report unadjusted limits, should Parsons ES consider them to be unacceptable?  Ms. Mullen responded that then they just don’t want to try.  She added that O’Brien & Gere will have to report unadjusted results for Dames & Moore.  Ms. Mullen stated that she would not accept an adjustment for dry weight; the adjustment is allowed by CLP.  

Denied.  Ms. Mullen indicated that CSSA is willing to pay to get the data right the first time.  She suggested that Parsons provide O’Brien & Gere with a program to make the conversion back to unadjusted limits.  If the lab can’t do it, and it is something that Parsons can do, Parsons should do it for them.  Parsons should notify AFCEE of additional costs.

 

8.2  Data Review, Validation, and Reporting Requirements for Definitive Data, p. 8‑2

The AFCEE QAPP requires that TICs be flagged with a “T.”  O’Brien & Gere requests a variance to allow the following:  TICs, when required, will be provided by hardcopy only on a separate result form from the “TCL” results.

Currently our instrument software does not electronically transfer this information into our LIMS system.

Not discussed during conference call.

None.

 

8.8, Hardcopy data reports for screening and definitive data, p. 8‑9

The AFCEE QAPP requires that hardcopy data reports conform to the formats identified in the referenced section.  O’Brien & Gere requests approval for in-house report forms be substituted for the required forms.  The results form that would be provided will include all required flags as specifed in the tables in section 7.2, Table 8.2-1, 8.2-2, 8.2-3, and 8.2-4.  See attached “Example Report Forms.”

We currently do not have the ability to provide the forms in the requested format.  However, all of the information is available on our in-house report forms.  See attached “Example Report Forms.”

Parsons ES will generate forms following AFCEE requirements in-house if O’Brien and Gere cannot do it.  Parsons ES will provide AFCEE with a cost estimate for additional effort associated with providing AFCEE forms.

Ms. Mirchandani asked if AFCEE 2.0 forms would be acceptable if the lab cannot do AFCEE 3.0 forms.  Ms. Mullen and Ms. Shively agreed that AFCEE 2.0 forms would be acceptable.

Parsons ES will get back to AFCEE on specific plan of action and associated costs.

Denied.  Either laboratory or Parsons ES must provide AFCEE forms.