[Home] [Master Table of Contents]
[SPCCP Index] [Appendix D Index]
Appendix D - Training Program
Photoionization/Flame Ionization Detectors
Objectives/Samples of Behavior:
Explain the operating principle of the PID & FID
Describe the calibration and operation of the PID & FID
Discuss the limitations of the PID & FID
Describe the advantages and applications of the PID & FID
A. Photoionization Detectors (PID)
Principles of operation
Ultraviolet (UV) lamps emit UV energy - photons - to the sample air
UV - edge of ionizing radiation
Molecules of contaminants in the air are ionized
Ions (charged particles) are collected at electrodes
Collection generates a current & subsequent meter reading "ppm"
UV lamps : 9.5 eV, 10.2 eV, 10.6 eV, 11.7 eV
UV energy must be equal to or greater than the E needed to remove the e in question
Energy required = ionization potential of the chemical
IPs found in NIOSH Pocket Guide
Using 10.6 eV lamp:
Substance |
IP eV |
Methylene chloride |
11.35 |
Ethylene |
10.52 |
n-Butane |
10.63 |
Acetone |
9.69 |
Methanol |
10.85 |
Isopropanol |
10.16 |
Ammonia |
10.15 |
Carbon monoxide |
14.01 |
Calibration - before each use / each day:
Zero:
"Zero gas"
Clean air
Span:
"Span gas" - isobutylene
Flush bag, attach tubing from instrument
Set instrument to span gas conc
Operation
Turn on, warm up
Place probe in suspect area
Can set alarm to various levels
Can run continuously or log data - e.g. record peak conc every 15 sec
Units:
ppm of span gas used - optimum conditions
ppm - cal gas equivalents or just "PID units"
Limitations
Water vapor, IP 12.59 - scatters, absorbs UV
Non-ionizable gases & vapors - blocking
Hi conc - inaccurate readings, e.g. drumhead space
Does not distinguish between substances - screens
Groups of chemical vapors: total ionizations
Can't "see" some chemicals - IP
Relative response - cal gas 100%; different gas ~75%, 110%, etc (ppm - cal gas equivalents)
Advantages / applications:
Wide variety of substances - good for fast screening
Detects both organic & inorganic
Sensitive to low conc
Helps in protective equipment selection
General IH work, IRP sites, emergency response, confined spaces, soil gas monitoring, etc.
B. Flame Ionization Detectors (FID)
Principles of operation:
Hydrogen gas is used to produce a flame in the detector
Flame is used to ionize organic molecules
- Will not pick up CO2, water vapor, etc.
Ions collected at electrodes & produce current / meter reading as with PID
IP not an issue - essentially all organics detected (IP 15.4 or less)
Calibration - before each use / each day:
Zero: zero gas or clean ambient air
Span: span gas - methane
Operation:
Insert external hydrogen fuel tank
Turn on, warm up (self-test - diagnostics)
Start pump, light hydrogen gas flame
Place probe in suspect areas
Can also alarm, data log
"FID units" or "ppm - cal gas equivalent"
Limitations
Can only read organic vapors / gases
Cannot distinguish between organics - get total organics
Needs external fuel source
Flame may extinguish at high vapor conc (or O2)
Relative response - e.g. most substances do not produce a reading of 100 ppm when their actual conc is 100 ppm
Substance |
Relative response, % |
Methane |
100 |
Hexane |
75 |
Vinyl chloride |
35 |
Benzene |
150 |
Acetylene |
225 |
MIBK |
100 |
Methanol |
12 |
Toluene |
110 |
Advantages / applications
Wide variety - detects more organics than PID
IP not a problem
Sensitive to low conc
Often used with gas chromatograph (GC) to help ID chemicals
Applications as with PID
Tremendous sampling variety when PID & FID used together