Measurement Parameter |
Control Checks |
Control Limits |
Corrective Actions* |
pH |
Measure buffer pH at least following every other measurement. |
Buffer measurements within 0.2 units of actual values. |
Recalibrate pH meter; check batteries and probe condition. |
Electrical Conductance |
Measure standard daily. |
±10 percent of actual value |
Replace or replatinize probe; verify that meter zeros and redlines properly; check batteries. |
Temperature |
Check measurement. |
±1șC |
Replace thermometer or correct temperature readings. |
Volatile organics (PID) |
Measure standard gas at least daily. |
±5 percent |
Replace dust filter; clean lamp; check battery. |
Explosive gases |
Measure calibration standard daily; test periodically in known explosive environment (gas tank) and check zero in clean environment. |
Adjust meter to read exact standard value. Control limit for fresh air is ±1 percent. |
Recalibrate meter; check battery; replace sensor. |
Water level (water level indicator) |
Measure weekly against tape measure when in regular use. |
±0.02 feet |
Replace meter tape. |
* Required if control limits not achieved. |